For measuring the air concentration of chemicals including Acrolein, Aldehyde screening, Formaldehyde, Furfural, Phosgene, and Valeraldehyde.
Description
These tubes are used to measure the average concentration of potentially harmful vapors in the air over a defined sample period. To use, the tube is connected to a small vacuum pump that draws air through the tube at a precise flow rate over a period of time. After the sample is taken, the tube is transported to a laboratory where they can remove the adsorbent material and perform analysis by solvent extraction and gas chromatography (GC) to determine the quantity of chemical collected, from which a time-weighted average (TWA) concentration can be calculated.
Specifications
Sorbent Material |
XAD-2® treated with 2-Hydroxymethyl Piperidine |
Flow Resistance |
≤ 10 kPa at flows up to 500cc/min |
Operating Temperature Range |
10 to 40o C between 20-80% RH |
Blank Tube Background |
≤ NIOSH LOD for the target vapor |
Standards Compliance |
ISO 22065 |
Product # |
Tube OD |
Tube Length |
Primary Sorbent Weight |
Backup Sorbent Weight |
ZST-085 |
6mm |
110mm |
150mg |
75mg |
ZST-086 |
6mm |
110mm |
120mg |
60mg |
ZST-087 |
8mm |
110mm |
450mg |
225mg |
Additional Information Available
Zefon XAD-2 Sorbent Tubes treated with 2-Hydroxymethyl Piperidine Operating Instructions