Nanoparticle Respiratory Deposition (NRD) Sampler User Manual

Download PDF

LA03061 Rev3

nanoparticle nrd sampler

Buy:

Nanoparticle (NRD) Sampler w/ 2" Foam Filter

Nanoparticle (NRD) Sampler w/ Nylon Filters

 

The Zefon Personal Nanoparticle Respiratory Deposition (NRD) samplers are designed to size-selectively collect nanoparticles with an efficiency that matches their deposition in the respiratory tract system. Their lightweight, disposable design allows them to be worn easily for personal exposure monitoring while providing the convenience of a preloaded ready to use the product.

Product Specifications

Housing Material: Carbon Filled Polypropylene

Filter Material (diffusion stage): Nylon or Foam (depending on model)

Collection Efficiency: <300nm with an efficiency designed to match the efficiency at which particles deposit in the respiratory tract

Cut-Point (cyclone stage): 4µm

Cut-Point (impactor stage): 300nm

Flow Rate: 2.5 LPM

 

Sampling Protocol

1. Remove the black plug from the NRD sampler inlet and insert the 25mm Aluminum cyclone.

remove black plug from nrd sampler

2. Place the NRD/Cyclone assembly into the cyclone holder.

place nrd/cyclone assembly into cyclone holder

3. Remove red outlet plug and pull the top retainer over the outlet of the NRD sampler.

remove red outlet plug

 

4. Connect the tubing adapter to the outlet of the NRD cassette.

connect tubing adapter

5. Check to make sure the cyclone is oriented correctly by making sure the locating nub is seated as shown.

check that cyclone is oriented correctly

6. Sample using a calibrated personal sampling pump at a flow rate of 2.5 LPM.

nrd completed assembly

Post Sampling

1. After sampling, remove the cyclone and re-install the black plug over the inlet. Make sure to push the plug all the way in for a tight seal.

2. Insert the red outlet plug.

3. Send to laboratory for analysis. (The Zefon NRD sampler is a new concept for personal nanoparticle exposure monitoring. As such, not all laboratories have defined an analysis protocol for the Zefon NRD sampler. Please contact your laboratory before sampling to discuss analysis options.)

 

Laboratory Analysisnrd lab analysis

1. In the laboratory, carefully open filter cassette at the diffusion stage.

2. Remove the series of diffusion filters or foam, digest, and process via chemical analysis.

3. The impaction stage is not designed to be analyzed and may be discarded.

 

nanoparticle respiratory deposition accessories chart

How it works

3 Stages of particle removal eliminates unwanted particles

Stage 3: Diffusion Stage

Air passes through a diffusion stage consisting of a series of nylon filters or a 2” section of foam, collecting particles <300nm with an efficiency designed to match the efficiency at which particles deposit in the respiratory tract.

Stage 2: Impaction Stage

Air passes through an impaction stage where particles larger than 300nm are removed using an impaction plate.

Stage 1: Respirable Cyclone (sold separately)

A respirable aluminum cyclone (Zefon #ZA0075) at a flow rate of 2.5 LPM is used to remove particles larger than 4µm in size.

znrd020znrd010 disassembled

Equipment Required

  • NRD Sampler
  • Personal sampling pump capable of a 2.5LPM flow rate at 15” H2 O backpressure
  • Calibration device for the sampling pump
  • A 25mm aluminum cyclone (Zefon model ZA0075 or equivalent)
  • A 25mm cyclone/cassette holder (Zefon model ZA0061 or equivalent)
  • An aluminum cyclone calibration adapter or calibration jar (Zefon model ZA0011 or equivalent)